Brierley, C.A., Benson, P.E. orcid.org/0000-0003-0865-962X and Sandler, J. (2017) How accurate are TheraMon® microsensors at measuring intraoral wear-time? Recorded vs. actual wear times in five volunteers. Journal of Orthodontics , 44 (4). pp. 241-248. ISSN 1465-3125
Abstract
BACKGROUND: The TheraMon® microsensor is the most recent device developed to measure the wear-time of removable appliances. The accuracy has not been validated intraorally.
OBJECTIVES: To determine 1) if the TheraMon® microsensor accurately records time when fixed intraorally, and 2) the effect of the intraoral location on the recorded time.
METHODS: A prospective pilot study, using a convenience sample, was carried out in a UK hospital orthodontic department. Five non-patient volunteers wore microsensors positioned palatal to an upper molar, and buccal to a lower molar for 7 days. Differences between actual amount of wear and the wear recorded by each device were calculated. Differences between sites were also examined.
RESULTS: The mean daily wear-time recorded by the upper and lower microsensors combined was 23 hours (95% CI 22.6-23.4), which is a mean under-recording of 4% (CI 2.5-5.8%). The maximum daily under-reporting of wear times was 5.5 hours. Microsensors in the lower buccal sulcus recorded wear-times that were closer to actual wear-times.
CONCLUSIONS: Assumptions made by the TheraMon® microsensors software lead to under-reporting of intraoral wear-time, particularly when placed palatally. These discrepancies could be significant in both clinical practice and research. Adjustment of the microsensor software parameters would improve accuracy, irrespective of the intraoral location.
Metadata
| Item Type: | Article |
|---|---|
| Authors/Creators: |
|
| Copyright, Publisher and Additional Information: | © 2017 Taylor & Francis. This is an author produced version of a paper subsequently published in Journal of Orthodontics. Uploaded in accordance with the publisher's self-archiving policy. |
| Keywords: | Microsensor; removable appliance; wear-time |
| Dates: |
|
| Institution: | The University of Sheffield |
| Academic Units: | The University of Sheffield > Faculty of Medicine, Dentistry and Health (Sheffield) > School of Clinical Dentistry (Sheffield) |
| Depositing User: | Symplectic Sheffield |
| Date Deposited: | 18 Sep 2017 10:03 |
| Last Modified: | 03 Nov 2023 16:41 |
| Status: | Published |
| Publisher: | Taylor & Francis |
| Refereed: | Yes |
| Identification Number: | 10.1080/14653125.2017.1365220 |