Culmer, PR, Brooks, PC, Strauss, DN et al. (4 more authors) (2014) An Instrumented Walking-Aid to Assess and Retrain Gait. IEEE Transactions on Mechatronics, 19 (1). 141 - 148. ISSN 1083-4435
Abstract
An instrumented walking-aid, the iWA system, has been developed to measure kinematic and kinetic properties of walking aid (WA) use and deliver feedback to improve gait. The clinical requirements, technical specification and design of the system are developed through clinical collaboration. The development of the system is described, including hardware components and data analysis used to process the measured data for assessment. The system measurements are validated under controlled laboratory conditions. The iWA system is evaluated in a typical UK clinical environment by a participant in a rehabilitation session. The resultant data successfully capture the quality of the participant’s walking aid use and agree with clinical opinion, supporting the efficacy of this approach.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © 2012 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other users, including reprinting/ republishing this material for advertising or promotional purposes, creating new collective works for resale or redistribution to servers or lists, or reuse of any copyrighted components of this work in other works |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Engineering & Physical Sciences (Leeds) > School of Mechanical Engineering (Leeds) |
Depositing User: | Symplectic Publications |
Date Deposited: | 14 Jul 2015 13:37 |
Last Modified: | 17 Jan 2018 18:13 |
Published Version: | http://dx.doi.org/10.1109/TMECH.2012.2223227 |
Status: | Published |
Publisher: | IEEE |
Identification Number: | 10.1109/TMECH.2012.2223227 |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:88038 |