Dahabreh, Z, Phillips, HK, Stewart, T et al. (1 more author) (2014) The effect of application time of two types of bone cement on the cement-bone interface strength. European Journal of Orthopaedic Surgery & Traumatology. ISSN 1633-8065
Abstract
The aim of this study was to investigate whether the application time of bone cement would have an effect on the cement-bone interface strength in two types of commercially available bone cements. CMW1 Radiopaque(®) (CMW1) and SmartSetHV(®) (SmartSet) were applied to bovine cancellous bone specimens at 2 and at 4 min. Specimens were loaded to failure and the shear strength of the cement-bone interface was calculated. The mean shear strength (±standard deviation) of the cement-bone interface was 2.79 ± 1.29 MPa for CMW1 applied at 2 min; 1.35 ± 0.89 MPa for CMW1 applied at 4 min; 2.93 ± 1.21 MPa for SmartSet applied at 2 min and 3.00 ± 1.11 MPa for SmartSet applied at 4 min. Compared to all other groups, the cement-bone interface strength was significantly lower when CMW1 was applied to the bone specimens at 4 min (p < 0.05). There was no significant difference in the cement-bone interface strength when SmartSet was applied to bone at 2 and at 4 min. Under these testing conditions, the cement-bone interface strength was not affected by the time of application of SmartSet to bone. However, it was significantly lower when CMW1 was applied to bone at 4 min.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © 2014, Springer. Reproduced in accordance with the publisher's self-archiving policy. The final publication is available in European Journal of Orthopaedic Surgery & Traumatology at Springer via DIO10.1007/s00590-014-1522-0 |
Keywords: | Bone cement; aseptic loosening; viscosity; total knee arthroplasty; polymethylmethacrylate |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Engineering & Physical Sciences (Leeds) > School of Mechanical Engineering (Leeds) > Institute of Medical and Biological Engineering (iMBE) (Leeds) |
Depositing User: | Symplectic Publications |
Date Deposited: | 12 Dec 2014 16:32 |
Last Modified: | 20 Apr 2015 18:27 |
Published Version: | http://dx.doi.org/10.1007/s00590-014-1522-0 |
Status: | Published |
Publisher: | Springer Paris |
Identification Number: | 10.1007/s00590-014-1522-0 |
Related URLs: | |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:81835 |
Download
Filename: Dahabreh Manuscript as submitted.pdf