Li, J, Redmond, AC, Jin, Z et al. (3 more authors) (2014) Hip contact forces in asymptomatic total hip replacement patients differ from normal healthy individuals: Implications for preclinical testing. Clinical Biomechanics, 29 (7). 747 - 751. ISSN 0268-0033
Abstract
Background: Preclinical durability testing of hip replacement implants is standardised by ISO-14242-1 (2002) which is based on historical inverse dynamics analysis using data obtained from a small sample of normal healthy individuals. It has not been established whether loading cycles derived from normal healthy individuals are representative of loading cycles occurring in patients following total hip replacement. Methods: Hip joint kinematics and hip contact forces derived from multibody modelling of forces during normal walking were obtained for 15 asymptomatic total hip replacement patients and compared to 38 normal healthy individuals and to the ISO standard for pre-clinical testing. Findings: Hip kinematics in the total hip replacement patients were comparable to the ISO data and the hip contact force in the normal healthy group was also comparable to the ISO cycles. Hip contact forces derived from the asymptomatic total hip replacement patients were comparable for the first part of the stance period but exhibited 30% lower peak loads at toe-off. Interpretation: Although the ISO standard provides a representative kinematic cycle, the findings call into question whether the hip joint contact forces in the ISO standard are representative of those occurring in the joint following total hip replacement.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | (c) 2014 Published by Elsevier Ltd. Uploaded in accordance with the publisher's self-archiving policy. NOTICE: this is the author’s version of a work that was accepted for publication in Clinical Biomechanics. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Clinical Biomechanics, 29, 7, (2014) DOI 10.1016/j.clinbiomech.2014.06.005 |
Keywords: | Arthroplasty; Contact forces; Hip; Kinematics; Preclinical testing; Wear |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Engineering & Physical Sciences (Leeds) > School of Mechanical Engineering (Leeds) > Institute of Medical and Biological Engineering (iMBE) (Leeds) |
Depositing User: | Symplectic Publications |
Date Deposited: | 06 Oct 2014 15:11 |
Last Modified: | 17 Jan 2018 03:01 |
Published Version: | http://dx.doi.org/10.1016/j.clinbiomech.2014.06.00... |
Status: | Published |
Publisher: | Elsevier |
Identification Number: | 10.1016/j.clinbiomech.2014.06.005 |
Related URLs: | |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:80549 |
Available Versions of this Item
- Hip contact forces in asymptomatic total hip replacement patients differ from normal healthy individuals: Implications for preclinical testing. (deposited 06 Oct 2014 15:11) [Currently Displayed]