Antochi, V.C., Baudis, L., Bollig, J. et al. (17 more authors) (2021) Improved quality tests of R11410-21 photomultiplier tubes for the XENONnT experiment. Journal of Instrumentation, 16 (08). P08033. ISSN 1748-0221
Abstract
Photomultiplier tubes (PMTs) are often used in low-background particle physics experiments, which rely on an excellent response to single-photon signals and stable long-term operation. In particular, the Hamamatsu R11410 model is the light sensor of choice for liquid xenon dark matter experiments, including XENONnT. The same PMT model was also used for the predecessor, XENON1T, where issues affecting its long-term operation were observed. Here, we report on an improved PMT testing procedure which ensures optimal performance in XENONnT. Using both new and upgraded facilities, we tested 368 new PMTs in a cryogenic xenon environment. We developed new tests targeted at the detection of light emission and the degradation of the PMT vacuum through small leaks, which can lead to spurious signals known as afterpulses, both of which were observed in XENON1T. We exclude the use of 26 of the 368 tested PMTs and categorise the remainder according to their performance. Given that we have improved the testing procedure, yet we rejected fewer PMTs, we expect significantly better PMT performance in XENONnT.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © 2021 The Author(s). Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence (http://creativecommons.org/licenses/by/4.0). Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. |
Keywords: | Photon detectors for UV, visible and IR photons (vacuum); Photon detectors for UV, visible and IR photons (vacuum) photomultipliers, HPDs, others); Dark Matter detectors (WIMPs, axions, etc.); Noble liquid detectors (scintillation, ionization, double-phase) |
Dates: |
|
Institution: | The University of Sheffield |
Academic Units: | The University of Sheffield > Faculty of Science (Sheffield) > School of Mathematical and Physical Sciences |
Funding Information: | Funder Grant number Swiss National Science Foundation 188716 |
Depositing User: | Symplectic Sheffield |
Date Deposited: | 22 May 2025 11:51 |
Last Modified: | 22 May 2025 11:51 |
Status: | Published |
Publisher: | IOP Publishing |
Refereed: | Yes |
Identification Number: | 10.1088/1748-0221/16/08/p08033 |
Related URLs: | |