Pippard, B.J., Neal, M.A., Maunder, A.M. et al. (8 more authors) (2021) Reproducibility of 19F‐MR ventilation imaging in healthy volunteers. Magnetic Resonance in Medicine, 85 (6). pp. 3343-3352. ISSN 0740-3194
Abstract
Purpose
To assess the reproducibility of percentage ventilated lung volume (%VV) measurements in healthy volunteers acquired by fluorine (19F)‐MRI of inhaled perfluoropropane, implemented at two research sites.
Methods
In this prospective, ethically approved study, 40 healthy participants were recruited (May 2018‐June 2019) to one of two research sites. Participants underwent a single MRI scan session on a 3T scanner, involving periodic inhalation of a 79% perfluoropropane/21% oxygen gas mixture. Each gas inhalation session lasted about 30 seconds, consisting of three deep breaths of gas followed by a breath‐hold. Four 19F‐MR ventilation images were acquired per participant, each separated by approximately 6 minutes. The value of %VV was determined by registering separately acquired 1H images to ventilation images before semi‐automated image segmentation, performed independently by two observers. Reproducibility of %VV measurements was assessed by components of variance, intraclass correlation coefficients, coefficients of variation (CoV), and the Dice similarity coefficient.
Results
The MRI scans were well tolerated throughout, with no adverse events. There was a high degree of consistency in %VV measurements for each participant (CoVobserver1 = 0.43%; CoVobserver2 = 0.63%), with overall precision of %VV measurements determined to be within ± 1.7% (95% confidence interval). Interobserver agreement in %VV measurements revealed a high mean Dice similarity coefficient (SD) of 0.97 (0.02), with only minor discrepancies between observers.
Conclusion
We demonstrate good reproducibility of %VV measurements in a group of healthy participants using 19F‐MRI of inhaled perfluoropropane. Our methods have been successfully implemented across two different study sites, supporting the feasibility of performing larger multicenter clinical studies.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © 2021 The Authors. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited (http://creativecommons.org/licenses/by/4.0/). |
Keywords: | 19F-MRI; lung; perfluoropropane; reproducibility; ventilation; Adult; Aged; Contrast Media; Female; Fluorine; Fluorocarbons; Humans; Lung; Lung Volume Measurements; Magnetic Resonance Imaging; Male; Middle Aged; Observer Variation; Prospective Studies; Reproducibility of Results; Young Adult |
Dates: |
|
Institution: | The University of Sheffield |
Academic Units: | The University of Sheffield > Sheffield Teaching Hospitals |
Funding Information: | Funder Grant number Medical Research Council MR/M008894/1; MR/N018915/1 |
Depositing User: | Symplectic Sheffield |
Date Deposited: | 25 Mar 2021 08:53 |
Last Modified: | 25 Mar 2021 08:53 |
Status: | Published |
Publisher: | Wiley |
Refereed: | Yes |
Identification Number: | 10.1002/mrm.28660 |