Winiberg, FAF, Smith, SC, Bejan, I et al. (6 more authors) (2015) Pressure-dependent calibration of the OH and HO2 channels of a FAGE HOx instrument using the Highly Instrumented Reactor for Atmospheric Chemistry (HIRAC). Atmospheric Measurement Techniques, 8 (2). pp. 523-540. ISSN 1867-8610
Abstract
The calibration of field instruments used to measure concentrations of OH and HO₂ worldwide has traditionally relied on a single method utilising the photolysis of water vapour in air in a flow tube at atmospheric pressure. Here the calibration of two FAGE (fluorescence assay by gaseous expansion) apparatuses designed for HOx (OH and HO₂) measurements have been investigated as a function of external pressure using two different laser systems. The conventional method of generating known concentrations of HOx from H₂O vapour photolysis in a turbulent flow tube impinging just outside the FAGE sample inlet has been used to study instrument sensitivity as a function of internal fluorescence cell pressure (1.8–3.8 mbar). An increase in the calibration constants CHO and CHO₂ with pressure was observed, and an empirical linear regression of the data was used to describe the trends, with ΔCHO = (17 ± 11) % and ΔCHO₂ = (31.6 ± 4.4)% increase per millibar air (uncertainties quoted to 2σ). Presented here are the first direct measurements of the FAGE calibration constants as a function of external pressure (440–1000 mbar) in a controlled environment using the University of Leeds HIRAC chamber (Highly Instrumented Reactor for Atmospheric Chemistry). Two methods were used: the temporal decay of hydrocarbons for calibration of OH, and the kinetics of the second-order recombination of HO₂ for HO₂ calibrations. Over comparable conditions for the FAGE cell, the two alternative methods are in good agreement with the conventional method, with the average ratio of calibration factors (conventional : alternative) across the entire pressure range, COH(conv)/COH(alt) = 1.19 ± 0.26 and CHO₂ (conv)/CHO₂ (alt) = 0.96 ± 0.18 (2σ). These alternative calibration methods currently have comparable systematic uncertainties to the conventional method: ~ 28% and ~ 41% for the alternative OH and HO₂ calibration methods respectively compared to 35% for the H₂O vapour photolysis method; ways in which these can be reduced in the future are discussed. The good agreement between the very different methods of calibration leads to increased confidence in HOx field measurements and particularly in aircraft-based HOx measurements, where there are substantial variations in external pressure, and assumptions are made regarding loss rates on inlets as a function of pressure.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © Author(s) 2015. This work is distributed under the Creative Commons Attribution 3.0 License. |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Engineering & Physical Sciences (Leeds) > School of Chemistry (Leeds) > Physical Chemistry (Leeds) |
Depositing User: | Symplectic Publications |
Date Deposited: | 16 Apr 2021 11:54 |
Last Modified: | 25 Jun 2023 22:35 |
Status: | Published |
Publisher: | Copernicus Publications |
Identification Number: | 10.5194/amt-8-523-2015 |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:171243 |