Clarke, EL, Munnings, C, Williams, B et al. (2 more authors) (2020) Display evaluation for primary diagnosis using digital pathology. Journal of Medical Imaging, 7 (02). p. 1. ISSN 2329-4302
Abstract
Purpose: As pathology departments around the world contemplate digital microscopy for primary diagnosis, making an informed choice regarding display procurement is very challenging in the absence of defined minimum standards. In order to help inform the decision, we aimed to conduct an evaluation of displays with a range of technical specifications and sizes.
Approach: We invited histopathologists within our institution to take part in a survey evaluation of eight short-listed displays. Pathologists reviewed a single haematoxylin and eosin whole slide image of a benign nevus on each display and gave a single score to indicate their preference in terms of image quality and size of the display.
Results: Thirty-four pathologists took part in the display evaluation experiment. The preferred display was the largest and had the highest technical specifications (11.8-MP resolution, 2100 cd / m2 maximum luminance). The least preferred display had the lowest technical specifications (2.3-MP resolution, 300 cd / m2 maximum luminance). A trend was observed toward an increased preference for displays with increased luminance and resolution.
Conclusions: This experiment demonstrates a preference for large medical-grade displays with the high luminance and high resolution. As cost becomes implicated in procurement, significantly less expensive medical-grade displays with slightly lower technical specifications may be the most cost-effective option.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © 2020 Society of Photo-Optical Instrumentation Engineers. This is an author produced version of a paper published in Journal of Medical Imaging. Uploaded in accordance with the publisher's self-archiving policy. |
Keywords: | digital pathology; display; monitor; whole slide image |
Dates: |
|
Institution: | The University of Leeds |
Depositing User: | Symplectic Publications |
Date Deposited: | 21 May 2020 12:17 |
Last Modified: | 21 May 2020 12:18 |
Status: | Published |
Publisher: | Society of Photo-optical Instrumentation Engineers |
Identification Number: | 10.1117/1.jmi.7.2.027501 |
Related URLs: | |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:161046 |