Al Musa, T, Uddin, A, Loveday, C et al. (14 more authors) (2019) Silent cerebral infarction and cognitive function following TAVI: an observational two-centre UK comparison of the first-generation CoreValve and second-generation Lotus valve. BMJ Open, 9 (1). e022329. ISSN 2044-6055
Abstract
Objective: To compare the incidence of silent cerebral infarction and impact on cognitive function following transcatheter aortic valve implantation (TAVI) with the first-generation CoreValve (Medtronic, Minneapolis, Minnesota, USA) and second-generation Lotus valve (Boston Scientific, Natick Massachusetts, USA).
Design: A prospective observational study comprising a 1.5 T cerebral MRI scan, performed preoperatively and immediately following TAVI, and neurocognitive assessments performed at baseline, 30 days and 1 year follow-up.
Setting: University hospitals of Leeds and Leicester, UK.
Patients: 66 (80.6±8.0 years, 47% male) patients with high-risk severe symptomatic aortic stenosis recruited between April 2012 and May 2015.
Main outcome measures: Incidence of new cerebral microinfarction and objective decline in neurocognitive performance.
Results: All underwent cerebral MRI at baseline and immediately following TAVI, and 49 (25 Lotus, 24 CoreValve) completed neurocognitive assessments at baseline, 30 days and 1 year. There was a significantly greater incidence of new cerebral microinfarction observed following the Lotus TAVI (23 (79%) vs 22 (59%), p=0.025) with a greater number of new infarcts per patient (median 3.5 (IQR 7.0) vs 2.0 (IQR 3.0), p=0.002). The mean volume of infarcted cerebral tissue per patient was equivalent following the two prostheses (p=0.166). More patients suffered new anterior (14 (48%) vs 2 (5%), p=0.001) and vertebrobasilar (15 (52%) vs 7 (19%), p=0.005) lesions following Lotus. Lotus was associated with a decline in verbal memory and psychomotor speed at 30 days. However, performance longitudinally at 1 year was preserved in all neurocognitive domains.
Conclusions: There was a higher incidence of silent cerebral microinfarction and a greater number of lesions per patient following Lotus compared with CoreValve. However, there was no objective decline in neurocognitive function discernible at 1 year following TAVI with either prosthesis.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © Author(s) (or their employer(s)) 2019. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/. |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Medicine and Health (Leeds) > School of Medicine (Leeds) > Leeds Institute of Cardiovascular and Metabolic Medicine (LICAMM) > Biomedical Imaging Science Dept (Leeds) The University of Leeds > Faculty of Medicine and Health (Leeds) > School of Medicine (Leeds) > Leeds Institute of Cardiovascular and Metabolic Medicine (LICAMM) > Division of Biomedical Imaging (Leeds) |
Funding Information: | Funder Grant number British Heart Foundation PG/11/126/29321 |
Depositing User: | Symplectic Publications |
Date Deposited: | 09 Apr 2019 15:27 |
Last Modified: | 25 Jun 2023 21:46 |
Status: | Published |
Publisher: | BMJ Publishing Group |
Identification Number: | 10.1136/bmjopen-2018-022329 |