Wyatt, J, Hedley, S, Johnstone, E et al. (7 more authors) (2018) Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy. Physics and Imaging in Radiation Oncology, 6. pp. 31-38. ISSN 2405-6316
Abstract
Background and purpose: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors.
Materials and methods: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation.
Results: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was for all scanners and sequences. For the lateral offset scan of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%).
Conclusions: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | (c) 2018 Published by Elsevier B.V. on behalf of European Society of Radiotherapy & Oncology. This is an open access article under the CC BY-NC-ND license (https://creativecommons.org/licenses/BY-NC-ND/4.0/). |
Keywords: | Geometric distortion; Magnetic resonance imaging; Radiotherapy; MR-only radiotherapy planning |
Dates: |
|
Institution: | The University of Leeds |
Depositing User: | Symplectic Publications |
Date Deposited: | 27 Jun 2018 15:35 |
Last Modified: | 25 Jun 2023 21:24 |
Status: | Published |
Publisher: | Elsevier |
Identification Number: | 10.1016/j.phro.2018.04.005 |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:132557 |