Dawkins, A. and Bjerre, A. (2016) Do the near computerised and non-computerised crowded Kay picture tests produce the same measure of visual acuity? British and Irish Orthoptic Journal, 13. pp. 22-28. ISSN 1743-9868
Abstract
Aims: Apps have been developed to assess visual acuity (VA) on tablet computers. The aim of this study was to compare near VA scores using crowded Kay pictures on the iSight app and the printed crowded Kay picture test in amblyopic and typically developing children to determine whether the tests are clinically interchangeable.
Methods: Fifty-seven participants (34 typically developing and 23 amblyopic children) aged 3–9 years had their uniocular near VA measured using computerised crowded Kay pictures on the iSight app for the iPad and the printed near crowded Kay picture test. Data were analysed to determine whether there was a significant difference between the two tests. Bland–Altman plots were constructed to assess how well the tests agreed with each other.
Results: There was no significant difference between the two tests in all test conditions with the exception of the left eyes of typically developing children measured using the per line termination criteria ( p = 0.01). Bland–Altman analysis showed good agreement between the iSight app and near crowded Kay picture test.
Conclusions: The results of the study indicate that near Kay pictures on the iSight app are clinically interchangeable with the traditional printed Kay pictures. These results compare well with other published studies comparing computerised vision tests with their traditional counterparts.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © 2016 the Author(s). Available under a Creative Commons license http://creativecommons.org/licenses/by/3.0/. |
Dates: |
|
Institution: | The University of Sheffield |
Depositing User: | Symplectic Sheffield |
Date Deposited: | 21 Jun 2018 08:42 |
Last Modified: | 21 Jun 2018 08:42 |
Published Version: | http://doi.org/10.22599/bioj.98 |
Status: | Published |
Publisher: | British and Irish Orthoptic Society |
Refereed: | Yes |
Identification Number: | 10.22599/bioj.98 |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:132368 |