Hua, X orcid.org/0000-0001-7512-660X, Li, J, Wang, L et al. (3 more authors) (2015) The effect of cup outer sizes on the contact mechanics and cement fixation of cemented total hip replacements. MEDICAL ENGINEERING & PHYSICS, 37 (10). pp. 1008-1014. ISSN 1350-4533
Abstract
One important loosening mechanism of the cemented total hip arthroplasty is the mechanical overload at the bone-cement interface and consequent failure of the cement fixation. Clinical studies have revealed that the outer diameter of the acetabular component is a key factor in influencing aseptic loosening of the hip arthroplasty. The aim of the present study was to investigate the influence of the cup outer diameter on the contact mechanics and cement fixation of a cemented total hip replacement (THR) with different wear penetration depths and under different cup inclination angles using finite element (FE) method. A threedimensional FE model was developed based on a typical Charnley hip prosthesis. Two acetabular cup designs with outer diameters of 40 and 43 mm were modelled and the effect of cup outer diameter, penetration depth and cup inclination angle on the contact mechanics and cement fixation stresses in the cemented THR were studied. The results showed that for all penetration depths and cup inclination angles considered, the contact mechanics in terms of peak von Mises stress in the acetabular cup and peak contact pressure at the bearing surface for the two cup designs were similar (within 5%). However, the peak von Mises stress, the peak maximum principal stress and peak shear stress in the cement mantle at the bone-cement interface for the 43 mm diameter cup design were predicted to be lower compared to those for the 40 mm diameter cup design. The differences were predicted to be 15–19%, 15–22% and 18–20% respectively for different cup penetration depths and inclination angles, which compares to the clinical difference of aseptic loosening incidence of about 20% between the two cup designs
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | (c) 2014 IPEM. Published by Elsevier Ltd. This is an author produced version of a paper published in Medical Engineering and Physics. Uploaded in accordance with the publisher's self-archiving policy. |
Keywords: | Contact mechanics; Cement fixation; Cup outer diameter; Inclination angle; Penetration; Total hip replacement |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Engineering & Physical Sciences (Leeds) > School of Mechanical Engineering (Leeds) > Institute of Medical and Biological Engineering (iMBE) (Leeds) |
Depositing User: | Symplectic Publications |
Date Deposited: | 26 Apr 2016 09:44 |
Last Modified: | 12 Dec 2024 11:36 |
Published Version: | http://http/dx.doi.org/10.1016/j.medengphy.2015.08... |
Status: | Published |
Publisher: | Elsevier |
Identification Number: | 10.1016/j.medengphy.2015.08.003 |
Related URLs: | |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:97358 |