Bissell, MM orcid.org/0000-0002-1282-2859, Loudon, M, Hess, AT et al. (4 more authors) (2018) Differential flow improvements after valve replacements in bicuspid aortic valve disease: a cardiovascular magnetic resonance assessment. Journal of Cardiovascular Magnetic Resonance, 20 (1). 10. ISSN 1097-6647
Abstract
Background
Abnormal aortic flow patterns in bicuspid aortic valve disease (BAV) may be partly responsible for the associated aortic dilation. Aortic valve replacement (AVR) may normalize flow patterns and potentially slow the concomitant aortic dilation. We therefore sought to examine differences in flow patterns post AVR.
Methods
Ninety participants underwent 4D flow cardiovascular magnetic resonance: 30 BAV patients with prior AVR (11 mechanical, 10 bioprosthetic, 9 Ross procedure), 30 BAV patients with a native aortic valve and 30 healthy subjects.
Results
The majority of subjects with mechanical AVR or Ross showed normal flow pattern (73% and 67% respectively) with near normal rotational flow values (7.2 ± 3.9 and 10.6 ± 10.5 mm2/ms respectively vs 3.8 ± 3.1 mm2/s for healthy subjects; both p > 0.05); and reduced in-plane wall shear stress (0.19 ± 0.13 N/m2for mechanical AVR vs. 0.40 ± 0.28 N/m2 for native BAV, p < 0.05). In contrast, all subjects with a bioprosthetic AVR had abnormal flow patterns (mainly marked right-handed helical flow), with comparable rotational flow values to native BAV (20.7 ± 8.8 mm2/ms and 26.6 ± 16.6 mm2/ms respectively, p > 0.05), and a similar pattern for wall shear stress. Data before and after AVR (n = 16) supported these findings: mechanical AVR showed a significant reduction in rotational flow (30.4 ± 16.3 → 7.3 ± 4.1 mm2/ms; p < 0.05) and in-plane wall shear stress (0.47 ± 0.20 → 0.20 ± 0.13 N/m2; p < 0.05), whereas these parameters remained similar in the bioprosthetic AVR group.
Conclusions
Abnormal flow patterns in BAV disease tend to normalize after mechanical AVR or Ross procedure, in contrast to the remnant abnormal flow pattern after bioprosthetic AVR. This may in part explain different aortic growth rates post AVR in BAV observed in the literature, but requires confirmation in a prospective study.
Metadata
Item Type: | Article |
---|---|
Authors/Creators: |
|
Copyright, Publisher and Additional Information: | © The Author(s). 2018. Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
Keywords: | Magnetic resonance imaging; Aortic disease; Bicuspid aortic valve; Aortic valve replacement; 4D flow |
Dates: |
|
Institution: | The University of Leeds |
Academic Units: | The University of Leeds > Faculty of Medicine and Health (Leeds) > School of Medicine (Leeds) > Leeds Institute of Cardiovascular and Metabolic Medicine (LICAMM) > Biomedical Imaging Science Dept (Leeds) |
Depositing User: | Symplectic Publications |
Date Deposited: | 21 Nov 2018 15:00 |
Last Modified: | 21 Nov 2018 15:00 |
Status: | Published |
Publisher: | BMC |
Identification Number: | 10.1186/s12968-018-0431-5 |
Related URLs: | |
Open Archives Initiative ID (OAI ID): | oai:eprints.whiterose.ac.uk:138879 |